TRITON 6 | 12
Type 1 + 2 (Ultrapure / Pure Water) from tap water with 10L tank
(Type 1 + 3 (Ultrapure / RO Water) from tap water with 50L tank)
The compact lab water solution
The TRITON 6 | 12 is modular-designed and the
smallest all-in-one laboratory water solution
available in 6 or 12 l/h.
Each system meets and exceeds
Type 1 and Type 2 water directly from tap water.
TRITON 6 |12 systems are optionally upgradable
to adapt to your specific requirements and make
your work as pleasant and comfortable as possible.
TRITON 6 | 12
Type 1 + 2 (Ultrapure / Pure Water) from tap water with 10L tank
(Type 1 + 3 (Ultrapure / RO Water) from tap water with 50L tank)
The compact lab water solution
The TRITON 6 | 12 is modular-designed and the
smallest all-in-one laboratory water solution
available in 6 or 12 l/h.
Each system meets and exceeds
Type 1 and Type 2 water directly from tap water.
TRITON 6 |12 systems are optionally upgradable
to adapt to your specific requirements and make
your work as pleasant and comfortable as possible.
Due to clear icons the controller allows you to operate the system intuitively just like a smartphone.
Dispensing reports ensure a history of all withdrawals. Additionally limit violations and errors will be stored to provide quality traceability over time.
All relevant values are permanently supervised. Further to the general information like conductivity, resisitivity, temperature, production rate and operating pressure, the TRITON sytem reminds you early enough when a change of consumables is necessary or whenever a parameter is out of specification. This guarantees the correctness of all analysis.
Due to clear icons the controller allows you to operate the system intuitively just like a smartphone.
Dispensing reports ensure a history of all withdrawals. Additionally limit violations and errors will be stored to provide quality traceability over time.
All relevant values are permanently supervised. Further to the general information like conductivity, resisitivity, temperature, production rate and operating pressure, the TRITON sytem reminds you early enough when a change of consumables is necessary or whenever a parameter is out of specification. This guarantees the correctness of all analysis.
The compact design is focused on achieving a high level of functionality and efficiency in a small space. TRITON systems are light and easy to transport or move around.
The compact system can be easily integrated in the lab environment. The extremely small footprint take up less space and make the TRITON ideal for crowded or limited spaces. TRITON systems optimizing the use of available space in a lab.
Despite the compact design, TRITON systems are fully packed with all the necessary components that make life in the laboratory easier and safer. All variants are generally equipped with a pressure reducer, leakage sensor, pressure sensor and flow sensor, among other things.
The compact design is focused on achieving a high level of functionality and efficiency in a small space. TRITON systems are light and easy to transport or move around.
The compact system can be easily integrated in the lab environment. The extremely small footprint take up less space and make the TRITON ideal for crowded or limited spaces. TRITON systems optimizing the use of available space in a lab.
Despite the compact design, TRITON systems are fully packed with all the necessary components that make life in the laboratory easier and safer. All variants are generally equipped with a pressure reducer, leakage sensor, pressure sensor and flow sensor, among other things.
Highest quality parts and components ensure that ultrapure water is always available and avoid downtime in the lab.
TRITON systems provide ultrapure water consistently to the trace level with the combination of different technoliogies. You can fully rely on the water purity for your critical experiments. Needless to say that your TRITON device complies with USP and GLP requirements.
The 185 | 254 nm UV lamp, which can be integrated as an option, not only prevents microbiological growth, but also minimizes total organic carbon (TOC) in purified water. The low TOC level of < 2 ppb ensures the reproducibility and reliability of your lab results. To ensure reliable removal of endotoxins, microorganisms and particles as well as DNase and RNase, you can also add a built-in ultrafilter to your TRITON system. This ultrafilter makes the unit ideal for use in cell cultivation and your other special laboratory applications.
The combination of long running, elaborated consumables and the possibility to change all filters separately and only when its needed, the annual costs will be minimized.
Its qualified design, combined with our own quality assurance system is the key to the outstanding quality of the TRITON series.
A complete range of services, including validation support using calibrated equipment and a qualification workbook is available on request.
Highest quality parts and components ensure that ultrapure water is always available and avoid downtime in the lab.
TRITON systems provide ultrapure water consistently to the trace level with the combination of different technoliogies. You can fully rely on the water purity for your critical experiments. Needless to say that your TRITON device complies with USP and GLP requirements.
The 185 | 254 nm UV lamp, which can be integrated as an option, not only prevents microbiological growth, but also minimizes total organic carbon (TOC) in purified water. The low TOC level of < 2 ppb ensures the reproducibility and reliability of your lab results. To ensure reliable removal of endotoxins, microorganisms and particles as well as DNase and RNase, you can also add a built-in ultrafilter to your HALIOS system. This ultrafilter makes the unit ideal for use in cell cultivation and your other special laboratory applications.
The combination of long running, elaborated consumables and the possibility to change all filters separately and only when its needed, the annual costs will be minimized.
Its qualified design, combined with our own quality assurance system is the key to the outstanding quality of the TRITON series.
A complete range of services, including validation support using calibrated equipment and a qualification workbook is available on request.
Your pure water storage tank ensures that high-quality Type 2 water is always on hand and ready to use. Modern design features, a circulation and advanced venting filtration protect against stagnation and bacterial contamination.
Place the tank under or separately next to the production unit. Use your valuable lab space exactly as it is needed.
Upgrade the system with an external 30L, 60L or 100L tank to transform the system into a Type 1 (ultrapure) and Type 3 (RO) system.
Check and adjust the tank level in one percent steps to your specific needs. An integrated leak water detector would prevent from a possible malfunction of the unit. As all values are continously monitored you can rely 100% on your TRITON system which saves time and money.
Your pure water storage tank ensures that high-quality Type 2 water is always on hand and ready to use. Modern design features and a circulation protect against stagnation and bacterial contamination.
Place the tank under or separately next to the production unit. Use your valuable lab space exactly as it is needed.
Upgrade the system with an external 30L, 60L or 100L tank to transform the system into a Type 1 (ultrapure) and Type 3 (RO) system.
Check and adjust the tank level in one percent steps to your specific needs. An integrated leak water detector would prevent from a possible malfunction of the unit. As all values are continously monitored you can rely 100% on your HALIOS system which saves time and money.
Feed water requirements | |
---|---|
Feed water quality | Potable tap water |
Conductivity @ 25°C | < 1000µS/cm |
Total organic carbon (TOC) | < 1ppm |
Inlet pressure | 1 - 6bar |
Temperature | 5 - 35°C |
Ultrapure water specifications | (Type 1) |
---|---|
Resistivity at 25 °C | 18.2 MΩ∙cm |
Total organic carbon (TOC) | ≤ 2 ppb ≤ 10 ppb (without UV) |
Particles content | Particle-free (size > 0.22 µm) |
Bacteria | < 0.01 CFU/mL < 0.005 CFU/mL (with ultrafilter) |
Pyrogens (endotoxins) | < 0.001 EU/mL (with ultrafilter) |
RNases | < 0.003 ng/mL (with ultrafilter) |
DNases | < 0.4 pg/µL (with ultrafilter) |
Flow rate | adjustable up to 2 L/min |
Pure water specifications | (Type 2) |
---|---|
Resistivity at 25 °C | > 10 MΩ∙cm |
Total organic carbon (TOC) | ≤ 30 ppb |
Production rate at 10°C | 6 L/h or 12 L/h |
Production rate at 20°C | 8L/h or 15 L/h |
Dimensions
Feed Water Requirements
Feed water quality | Potable tap water |
---|---|
Conductivity @ 25°C | < 1000µS/cm |
Total organic carbon (TOC) | < 1ppm |
Inlet pressure | 1 - 6bar |
Temperature | 5 - 35°C |
Ultrapure water (Type 1) Specifications
Resistivity at 25 °C | 18.2 MΩ∙cm |
---|---|
Total organic carbon (TOC) | ≤ 2 ppb ≤ 10 ppb (without UV) |
Particles content | Particle-free (size > 0.22 µm) |
Bacteria | < 0.01 CFU/mL < 0.005 CFU/mL (with ultrafilter) |
Pyrogens (endotoxins) | < 0.001 EU/mL (with ultrafilter) |
RNases | < 0.003 ng/mL (with ultrafilter) |
DNases | < 0.4 pg/µL (with ultrafilter) |
Flow rate | adjustable up to 2 L/min |
Pure water (Type 2) Specifications
Resistivity at 25 °C | > 10 MΩ∙cm |
---|---|
Total organic carbon (TOC) | ≤ 30 ppb |
Production rate | 6 L/h or 12 L/h |
Dimensions